Technical data

Preventive Maintenance 3
Agilent N5508A Hardware Reference 41
Instrument Removal
To remove an instrument from the system, use one of the following procedures.
Required tools
#2 Phillips screwdriver
#2 POZIDRIV screwdriver
Half-Rack-Width instrument
To remove a half-width instrument from a system rack
1 Power off the system. For details, see the system installation
guide.
2 Remove the selected instruments power cord
from the power strip in the rack.
3 The instrument is attached to the half-rack width
instrument beside it; remove that instruments
power cord from the power strip also.
The instruments are secured together
by lock links at the front and rear. The
lock links at the rear attach with screws.
The lock links at the front hook together.
4 Remove the power cord and other cables from the
front and rear of both instruments.
Note the location of cables for
re-installation.
5 Remove the four corner screws on the front of the
rack panel that secures the instruments in place.
The screws are located near the corners
of the face of the instrument.
Use a #2 Phillips screwdriver.
6 Slide both instruments, as a single unit, out from
the front of the rack and set them on a secure, flat
surface.
7 Detach the lock links that secure the rear of the
instruments together by removing their screws.
Use a #2 POZIDRIV screwdriver.
See Figure 11 on page 42.
8 Carefully and at the same time, push one
instrument forward and pull the other back to
unhook the lock links that secure the front of the
instruments to each other.
9 Store the partner instrument and lock links
while the selected instrument is out of the rack.
Only install the instruments as a pair;
individual installation is not secure.