Technical data
Preventive Maintenance 3
Agilent N5508A Hardware Reference 37
Tab l e 13 Cleaning Supplies Available from Agilent
Product Part Number
Aero-Duster 8500-6460
Isopropyl alcohol 8500-5344
Lint-Free cloths 9310-0039
Small polyurethane swabs 9301-1243
WARNING
Cleaning connectors with alcohol should only be performed with the
instruments’ mains power cord disconnected, in a well ventilated area.
Connector cleaning should be accomplished with the minimum amount of
alcohol. Prior to connector reuse, be sure that all alcohol used has dried, and that
the area is free of fumes.
WARNING
If flammable cleaning materials are used, the material should not be stored, or
left open in the area of the equipment. Adequate ventilation should be assured to
prevent the combustion of fumes, or vapors.