Technical data
14 Agilent N5508A Hardware Reference
1 General Information
IF (OUTPUT)
The signal at this connector is the downconverter’s output.
Limits
• Nominal output level: 0 to +5 dBm (input signal ≥ –30 dBm)
• Maximum output level: +15 dBm
• Frequency (RF input 5 to 1500 MHz): 5 to 1500 MHz
• Frequency (RF input 5 to 26.5 GHz): 300 to 900 MHz
Characteristics
• Output impedance: 50 Ω
RF SIGNAL (INPUT)
This connector accepts a 120 MHz signal from a synthesized signal generator.
This signal is mixed with the internal microwave local oscillator. The result is a
tunable microwave signal with a 0.1 Hz resolution over most of the frequency
range.
Limits
• Maximum input level: +10 dBm
• Recommended input level: –10 to +5 dBm
• Frequency: 120 to 900 MHz
Characteristics
• Input impedance: 50 Ω
NOTE
The IF amplifiers frequency response starts rolling off above 1200 MHz. Avoid using IF
frequency between 1200 and 1500 MHz.