Technical Manual
Technical Manual  
ARGUS AM Floor System     
Cross Point Device Explorer  v1.2  Page 46 of 65 
  Noise level indicator 
The noise level indicator, located on the right of each scope view image, shows a value 
between  zero  and  one  hundred,  which  represents  the  amount  of  noise  that  is  being 
detected. A high noise level will result in reduced detection, which means that specified 
detection distances might not be achieved. To improve this situation, the noise source(s) 
must be located and removed when possible. 
  Extended configuration views 
When besides  the transceiver antenna  also receiver  antennas  are installed,  the  scope 
views of these receivers can also be viewed and the phase and anti-phase settings can be 
adjusted individually for each connected antenna. 
Figure 45: Scope view of transceiver and receiver antenna signals 
NOTE: The Auto threshold  setting is a global setting, which means  that it is  set  for all 
channels. 
  Triggering an alarm 
When the signal of tag or label is detected by any of the connected antennas, the control 
unit will trigger the tag detected alarm notification. An alarm is only triggered when both the 
SIGNAL and PHASE requirements are met. This means that the signal must be higher than 
the threshold level and a constant threshold degrees angle has been detected. If one of 
these requirements is not met, then the alarm will not be triggered. 
Situation 1 
SIGNAL too low and continuous PHASE detected; result: no alarm 
Toggle these buttons to view 
or hide the associated scope 
view(s) 










