User guide

6.2.5 Analog Input
AC coupled differential analog inputs.
Symbol Parameter Min Type Max Units
V
com
input common mode -4 6 V
V
p-p
Differential input Voltage -125 125 mV
Z
P
input impedance 100
analog offset -0,25 0.25 V
6.3 Information Required by DIN EN 61010-1
6.3.1 Manufacturer
The Ndigo5G is a product of:
cronologic GmbH & Co. KG
Jahnstraße 49
60318 Frankfurt
HRA 42869 beim Amtsgericht Frankfurt/M
VAT-ID: DE235184378
6.3.2 Intended Use and System Integration
The devices are not ready to use as delivered by cronologic. It requires the development of
specialized software to fulfill the application of the end user. The device is provided to sys-
tem integrators to be built into measurement systems that are distributed to end users. These
systems usually consist of a the Ndigo5G, a main board, a case, application software and pos-
sible additional electronics to attach the system to some type of detector. They might also be
integrated with the detector.
The Ndigo5G is designed to comply with DIN EN 61326-1 when operated on a PCI compliant
main board housed in a properly shielded enclosure. When operated in a closed standard
compliant PC enclosure the device does not pose any hazards as defined by EN 61010-1.
Radiated emissions, noise immunity and safety highly depend on the quality of the enclosure.
It is the responsibility of the system integrator to ensure that the assembled system is compliant
to applicable standards of the country that the system is operated in, especially with regards
to user safety and electromagnetic interference. Compliance was only tested for attached cables
shorter than 3m.
All power supplied to the system must be turned off before installing the board.
When handling the board, adequate measures have to be taken to protect the circuits against
electrostatic discharge (ESD).
cronologic GmbH & Co. KG 47 Ndigo5G User Guide