Specifications

SpO,values are also retained the same manner as
the BP and Temperature values.
Print
Selection of
this button causes the current data to be printed.
SpO, plethysmograph is displayed on the LCD,
the waveform pauses
for 2 minutes or until the Clear button is
selected.
SpO, monitoring and
alarms are suspended.
Clear
Selection of this button halts measurements and returns the
user to the Main menu.
Note: If the
LEDs
and
held for 2 minutes or until cleared by the user. The
LCD
shows:
Values Held
Note:
If the
printer is in “Auto”
mode, the Print button does not
appear as an option.
Notes
l If the Monitor is performing a Vitals determination, the
Vitals button cannot be selected.
l If a BP determination is in progress, the Vitals button
cannot be selected.
l A Vitals determination is canceled if the BP determination
is canceled.
. A Vitals determination can be canceled by pressing either
the AUTO
BP or START/STOP key.
l During the 2 minute freeze period,
SpO,
and predictive temperature determinations to
be monitored and recorded (depending on Monitor model).
When the Vitals determination is complete, a single “warble”
sounds and all patient data are displayed on the
-
Using the Menu System
Vitals Button (UK: All Obs)
Selection of this button initiates a BP determination while
allowing