Service manual

DINAMAP
Compact Vital Signs Monitor
76
Procedural & Error Alarm Code Table continued
Alarm
Code
LED
Display
LCD Display Audio Tone &
Volume
Effect of
Alarm
Silence
switch
Effect of
Clear via
Rotor
Probable Cause
No code
No
change
Blank
No
change
Blank
Low Battery
Flashing
battery icon
Low Battery -
System
Disabled
Printer - No
Paper
NIBP Range
error
3 beeps every
10 sec,
adjustable
volume.
Steady tone,
maximum
volume
High priority
alarm
Volume
adjustable
High priority
alarm
Volume
adjustable
2 minutes
silence
No effect
2 minute
silence
2 minute
silence
No effect
No effect
Clear
Clear
Replace or recharge battery. From
onset of alarm, 5 NIBP
measurements available. Beep
linearly increases rate as battery
discharges.
Replace or recharge battery. NIBP
measurement disabled.
Paper expired or printer door open.
NIBP algorithm returned value
outside specified accuracy rage.
Other:
N,P,E,I,S.
Blank Error code,
description
Steady tone,
maximum
volume
No effect No effect Internal system fault