Service manual
DINAMAP
™
Compact Vital Signs Monitor
75
6.5 Procedural & Error Alarm Code Table
Alarm
Code
LED
Display
LCD Display Audio Tone &
Volume
Effect of
Alarm
Silence
switch
Effect of
Clear
via
Rotor
Probable Cause
N99
N55
N44
N33
N00
No
change
N99-NIBP
FAILED
N55-
TIMEOUT:
PRESS
N44-
TIMEOUT:
TOTAL
N33-
TIMEOUT:
INFLATE
N00- OVER
PRESSURE
High priority
alarm. Volume
adjustable.
2 minutes
silence
Clear
Unable to make an NIBP determination
due to insufficient signal.
One cuff pressure for > 1 minute. Motion
artifact.
Determination time exceeds 2 minutes.
Motion artifact.
Inflation time > 40 seconds or air leak
detected.
Overpressure detected.
N12 Systolic
Display
‘N12’
N0 Display Steady tone,
maximum
volume
No effect No effect System pressure whilst idle 5 or
15mmHg above ambient for 30 seconds
for a neonate or an adult cuff
respectively
N13 Systolic
Display
‘N13’
No Display Steady tone,
maximum
volume
No effect No effect Pump on/off status determined by
measurement of pump current does not
concur with Main board software
expected status. Unit will power down
within 30 seconds of detected failure..
P55
P00
No
change
P55-SpO
2
: NO
SIGNAL
P00-SpO
2
: NO
SpO
2
SENSOR
High priority
alarm. Volume
adjustable.
2 minutes
silence
Clear
No or very low SpO
2
signal. Check or
reposition sensor.
SpO
2
sensor not connected. No sensor
code detected. Sensor failure.
E33
E11
E00
No
change
E33-Temp:
FAIL
E11-Temp:
FAIL
E00-Temp:
FAIL
High priority
alarm. Volume
adjustable.
2 minutes
silence
Clear
Clear
Clear
Temperature probe not connected or
inoperable.
Predictive temperature loss of tissue
contact > 30 sec.
Removal of probe before temperature
determination complete.
Predictive temperature determination
exceeds 60 sec.