Service manual
37
DINAMAP
™
COMPACT Calibration check or re-calibration
3.3.2 NIBP Calibration
The following section describes how to perform NIBP re-calibration on a DINAMAP
™
Compact Vital Signs
Monitor. Refer also to the previous section for details on checking of the DINAMAP
™
Compact calibration.
NIBP calibration is possible without disassembling the DINAMAP
™
Compact.
I. Set up the DINAMAP
™
Compact and calibration equipment as shown in Section 3.3.2.
II. Switch on the DINAMAP
™
Compact and allow 10 seconds to settle.
III. Verify that the calibration pressure measurement system reads 0 mmHg (0 mBar)
IV. If the DINAMAP
™
Compact displays the language choice menu, choose “English”, then “OK”.
Reboot the DINAMAP
™
Compact by turning monitor off, and then back on.
V. Select the “Mode” button from the main menu. The Mode menu will be displayed
VI. Select the “Service” button from the mode menu
VII. Enter the Service Mode PIN code, 2 2 1 3. The Service menu will be displayed.
VIII. Select “NIBP” from the service menu. The NIBP Service menu will be displayed.
IX. Select “Cal” from the NIBP service menu.
-- Note --
From this point, the timing is critical.
It is recommended to read through
the following steps before proceeding.