Service manual

37
DINAMAP
COMPACT Calibration check or re-calibration
3.3.2 NIBP Calibration
The following section describes how to perform NIBP re-calibration on a DINAMAP
Compact Vital Signs
Monitor. Refer also to the previous section for details on checking of the DINAMAP
Compact calibration.
NIBP calibration is possible without disassembling the DINAMAP
Compact.
I. Set up the DINAMAP
Compact and calibration equipment as shown in Section 3.3.2.
II. Switch on the DINAMAP
Compact and allow 10 seconds to settle.
III. Verify that the calibration pressure measurement system reads 0 mmHg (0 mBar)
IV. If the DINAMAP
Compact displays the language choice menu, choose English, then OK.
Reboot the DINAMAP
Compact by turning monitor off, and then back on.
V. Select the Modebutton from the main menu. The Mode menu will be displayed
VI. Select the Servicebutton from the mode menu
VII. Enter the Service Mode PIN code, 2 2 1 3. The Service menu will be displayed.
VIII. Select NIBPfrom the service menu. The NIBP Service menu will be displayed.
IX. Select Calfrom the NIBP service menu.
-- Note --
From this point, the timing is critical.
It is recommended to read through
the following steps before proceeding.