Service manual

36
3.3 Calibration - Checking & Re-calibrating
This section details the following procedures:
?? Checking NIBP calibration
?? Recalibrating NIBP
?? Checking NIBP overpressure detection (Pop Off)
?? Re-calibrating NIBP overpressure detection
?? Predictive temperature calibration checking
?? Re-calibration of predictive temperature
3.3.1 Checking of NIBP Calibration
The following section describes how to verify the NIBP calibration on a DINAMAP
Compact Vital Signs
Monitor. Refer also to the next section for details on re-calibration of a DINAMAP
Compact.
Checking of NIBP calibration is possible without disassembling the DINAMAP
Compact.
I. Set up the DINAMAP
Compact and calibration equipment as shown on page 38.
II. Switch on the DINAMAP
Compact and allow 10 seconds to settle.
III. Verify that the calibration pressure measurement system reads 0 mmHg (0 mBar)
IV. If the DINAMAP
Compact displays the language choice menu, choose English, then OK.
Reboot the DINAMAP
Compact by turning monitor off, and then back on.
V. Select the Modebutton from the main menu. The Mode menu will be displayed
VI. Select the Servicebutton from the mode menu
VII. Enter the Service Mode PIN code, 2 2 1 3. The Service menu will be displayed.
VIII. Select NIBPfrom the service menu. The NIBP Service menu will be displayed.
IX. Select Checkfrom the NIBP service menu.
X. The DINAMAP
Compact now displays pressures on the systolic and diastolic displays.
XI. Check calibration of the DINAMAP
Compact unit is within specification at 0mmHg and
200mmHg (267mBar) applied pressure using a reference pressure measurement system.
Display Meaning Units System Range
Systolic Pressure measured at PT1
(main pressure transducer, on
Main Board)
mmHg 0-300
Diastolic Pressure measured at PT2 (on
pump assembly)
mmHg 0-330
Pop Off will occur
before 330
Normal operation of the DINAMAP
Compact is not possible until powered off and back on.