Service manual
33
Cal
Selection of this button initiates the calibration procedure. See section 3.4.1 for a more detailed description
of using this mode.
Pop Off
Selection of this button initiates the Pop Off test. See section 3.4.3 for a more detailed description of using
this mode.
OK
Selection of this button returns the display to the Service Menu.
3.2.2.2 Temp
Selection of this button causes the Temperature service menu to appear.
C or F
Selection of this button allows the temperature measurement units to be toggled between Celsius and
Fahrenheit. A warning will appear to inform the user that this change will result in the loss of any existing
trend information.
Cal
Selection of this button causes the current date to be saved for the Info display. This should only be done
when the temperature calibration has been set as described in section 3.3.6.
3 Mins
Selection of this button sets the temperature mode to “3 minute predictive” i.e. the results obtained are
those which would be obtained by a mercury-in-glass thermometer after 3 minutes.
12 Mins
Selection of this button sets the temperature mode to “12 minute predictive” i.e. the results obtained are
those which would be obtained by a mercury-in-glass thermometer after 12 minutes.
OK
Selection of this button returns the display to the Service Menu.
3.2.2.3 Info
Selection of this button displays the last calibration date information for both the NIBP and Temperature
systems.
Selection of OK returns the display to the Service Menu.