Service manual

29
3.2 Service Modes
There are three different service modes, each offering different levels of access to the monitors set-up and
calibration. These modes can be accessed using the rotor control and display panel. The modes are:
1. The Clinician Menumode, allowing:
?? setting of temperature units,
?? checking NIBP and temperature calibration dates,
?? setting the alarms into Permanent Silencemode.
(This mode is described in the DINAMAP
Compact Monitor Operating Manual).
2. The Main Service Modeprovides the features allowed from the clinician menu (described
above) plus the following:
?? Checking NIBP calibration,
?? NIBP re-calibration,
?? Pop Off (overpressure detection) calibration,
?? Changing temperature algorithm,
?? Set temperature calibration date,
?? Configuration for use with Central Station,
?? Changing language of operation,
?? Monitor Set-up- specifying whether SpO
2
, printer, and temperature are fitted.
3. The Super Service Modeis used to allow the calibration of the Predictive Temperature circuitry and
Neonatal Overpressure checking procedure. Other menu options within this mode are for
manufacturing test purposes only.
Note : Certain menu functions may not be present or appearance may be slightly different than depicted in
the following section if unit is fitted with software earlier than 8613-V4.1.
To enter the service mode from the Main Menu, use the rotor control and LCD.
First select More
then Service