Service manual
12
PREDICTIVE TEMPERATURE ? 0.6
o
C, 36.1
- 39.4
o
C ± 1
o
F, 97
- 103
o
F
ACCURACY : UNSPECIFIED < 36.1
O
C, > 39.4
O
C
UNSPECIFIED < 97
O
F, > 103
O
F
DETERMINATION TIME: 30 seconds typical; 60 seconds maximum
SpO
2
(TS & S Model)
OXYGEN SATURATION
RANGE: 0 to 100%
SATURATION ACCURACY: 0% to 69%: unspecified
ADULT ACCURACY (70-100%)
Nellcor Puritan Bennett
™
Sensor Accuracy
OXICLIQ-P Pediatric sensor 2.5 digits
OXIBAND Pediatric/infant sensor 3.0 digits
DURA-Y ear clip 3.5 digits
OXISENSOR II D-20 Pediatric sensor 3.5 digits
OXICLIQ-N neonatal/adult sensor 2.5 digits
REFLECTANCE sensor 3.5 digits
DURASENSOR adult 3.0 digits
OXIBAND adult/neonatal sensor 3.0 digits
DURA-Y multisite sensor 3.0 digits
OXISENSOR R-15 adult nasal sensor 3.5 digits
OXISENSOR II D-25 adult sensor 2.0 digits
OXICLIQ-A adult sensor 2.5 digits
OXISENSOR II N-25 neonatal/adult sensor 2.0 digits
OXISENSOR II I-20 infant sensor 2.0 digits
OXISENSOR II D-25L adult sensor, long cable 2.0 digits
Neonatal Accuracy:
When sensors are used on neonatal subjects as recommended, the specified accuracy range is increased
by ?1 digit to account for the theoretical effect on oximeter measurements of fetal hemoglobin in neonatal
blood (e.g., N –25 accuracy on neonates is ?3, rather than ?2.)
Note: Refer to NELLCOR PURITAN BENNETT sensor specifications.
PULSE RATE DETERMINATION: 20 BPM - 250 BPM +/- 3 beats
SATURATION PITCH INDICATOR: Pitch changes with saturation.
Volume selectable from 0 (off) to 9.
WAVEFORMS: Pulse Plethysmograph waveform on LCD gain compensated
SENSOR DISCONNECT / The monitor will detect the attachment or disconnection
DISCONNECT FROM PATIENT: of a sensor from the patient within 5 seconds.
PULSE DETECTION: The monitor will detect a pulse or enter a no signal state within 15
seconds of being attached to the patient.