1.6.3 Connector Symbols
External Alarm Connector
External Comms Port Connector
Fuse
External Power
1.7 Electrical, Mechanical & Environmental Specifications
NIBP
CUFF PRESSURE RANGE: Adult / Pediatric: 0 mmHg to 290 mmHg
Neonate: 0 mmHg to 140 mmHg
DEFAULT TARGET Adult / Pediatric: 178 ± 15 mmHg
CUFF INFLATION: Neonate: 110 ± 15 mmHg
TARGET CUFF INFLATION Adult / Pediatric: 100 to 250 mmHg; 5 mmHg increments
ADJUSTMENT RANGE: Neonate: 100 to 140 mmHg; 5 mmHg increments
BLOOD PRESSURE
MEASUREMENT RANGE:
Systolic
mmHg
MAP mmHg Diastolic
mmHg
Adult / Pediatric: 30 - 245 15 - 215 10 - 195
Neonate: 40 - 140 30 - 115 20 - 100
BLOOD PRESSURE ACCURACY: Meets AAMI/ANSI SP-10 Electronic or Automated
Sphygmomanometer Standard for Accuracy. (AAMI/ANSI
standard :± 5 mmHg mean error; 8 mmHg standard deviation;
intra-arterial method.)
BLOOD PRESSURE 20 seconds to 45 seconds typical
DETERMINATION TIME: Adult: 120 seconds maximum,
Neonate: 85 seconds maximum.
PULSE RATE DETERMINATION: Adult / Pediatric: 30 - 200 BPM
Neonate: 30 - 200 BPM
PULSE RATE ACCURACY: ± 3.5 percent
OVERPRESSURE CUT-OFF: Adult / Pediatric 300 - 330 mmHg
Neonate 150 - 165 mmHg
Predictive Temperature Determination (TS & T models)
RANGE: Max. 42.2 ?Celsius 108.0° Fahrenheit
Min 31.1?Celsius 88.0° Fahrenheit
PROBE ACCURACY: ? 0.1
o
C ± 0.2
o
F