Specifications

9
• Place the Compact Monitor on a rigid, secure surface.
Monitor must only be used with mounting hardware,
poles, and stands recommended by Critikon. See
Appendix D.
• The weight of the accessory basket contents should
not exceed 6.6 lb (3 kg).
• Arrange the power cord, air hoses, and all cables
carefully so they do not constitute a hazard.
• Verify calibration of BP and TEMP (Models T and TS)
parameters (pulse oximeter does not require
calibration). Ensure that the Compact Monitor is
functioning properly before operating the Compact
Monitor.
• Do not immerse the Monitor in water. If the Monitor
is splashed with water or becomes wet, wipe it
immediately with a dry cloth.
• Do not gas sterilize or autoclave.
Notes
• Waveforms may be distorted and readings inaccurate
when electrosurgical cautery equipment is used while
monitoring with the Compact Monitor.
• The electromagnetic compatibility profile of the
Compact Monitor may change if accessories other than
those specified for use with the Compact Monitor are
used.
• Trend data are retained in the Compact Monitor when it
is turned off, except when the default is overridden by
selecting the Trend button under the Service menu.
Introduction
COMPACT MAN-REV B 1/12/99, 9:46 AM9