Specifications

87
Appendix G
Connection Details
Host Port Connector (rear panel)
Pin 15
Pin 1
Important! All host port signals are NON-ISOLATED and should
be connected to equipment conforming to IEC-601-1 ONLY.
Where isolation of data communication is required, Critikon part
number ILC232 should be used. If external alarm control is
required, Critikon part number 487208 (Isolated Remote Alarm
Cable Assembly) should ALWAYS be used. Please refer to the
Information Sheet included with the isolated remote alarm cable
for operational details.
Note: When using remote alarm, the Compact Monitor should be
considered the primary alarm source. The secondary alarm is
used for secondary purposes only.
Pin Number Function
1 Ground
2 Inverted TTL Transmit Data
3 Inverted TTL Receive Data
4 Fused +5 volts
5 No connection
6 No connection
7 Ground
8 Remote Alarm
9 RS232 Request to Send (RTS)
10 RS232 Clear to Send (CTS)
11 RS232 Transmit Data (TxD)
12 No connection
13 RS232 Receive Data (RxD)
14 No connection
15 No connection
COMPACT MAN-REV B 1/12/99, 9:47 AM87