Specifications
63
Neonatal Accuracy
When sensors are used on neonatal subjects as
recommended, the specified accuracy range is increased by
±1 digit to account for the theoretical effect on oximeter
measurements of fetal hemoglobin in neonatal blood (e.g.,
N-25 accuracy on neonates is ±3, rather than ±2.)
Note: Refer to NELLCOR’s sensor specifications.
Pulse Rate Range 20 to 250 beats/min
Accuracy ±3 beats/min
Audible Indicator Pitch changes with
saturation; volume
selectable from 0 (off) to 9
Waveforms Pulse plethysmograph
waveform on LCD gain
compensated
Sensor Disconnect/ Monitor will detect
Disconnect From Patient attachment or
disconnection of sensor
from patient within 5 s
Pulse Detection Monitor will detect pulse
or enter no signal state
within 15 s of being
attached to patient
Loss of Pulse Monitor will detect loss of
pulse from patient and enter
no signal state within 10 s
Appendix A