Specifications
59
Using the Menu System
Set BP. Selection of this button displays the message: Restore BP
mode on power-up? Selection of Yes restores the BP mode to
previous user-selected mode after power-off and returns the Monitor
to the More... menu. Selection of No restores the BP mode to the
default setting of Manual after power-off and returns the Monitor to
the More... menu. Selection of Cancel returns the Monitor to the
More... menu.
Note: The Monitor must be turned off and then on again to activate
this feature.
Alarms. Selection of this button displays the message: Enter alarm
configuration mode? Selection of No returns the Monitor to the
More... menu. Selection of Yes brings up the Alarms menu. Selection
of Reset changes all alarm limits back to the factory defaults and
returns the Monitor to the More... menu. Selection of Save
permanently saves the user-selected alarm limits and returns the
Monitor to the More... menu. Selection of Cancel returns the Monitor
to the More... menu.
Note: The Monitor must be turned off and then on again to activate
this feature.
Reset
SYS DIA MAP BPM SpO
2
Save
Cancel
180 130 140 160
OFF
30 15 50 40 90
Main. Selection of this button returns the user to the Main menu.
Silence. Selection of this button will cause all alarms except the
FAILSAFE alarm to be muted. A confirmation menu will appear in
Area 2 of the LED. Selection of either Yes or No returns the user to
the Clinician mode menu. If silence is confirmed, the Alarm Silence
button (11) illuminates and alarms are permanently muted. If silence
is not confirmed, the alarm will be audible.
Caution: Alarms will be muted until either the Monitor is switched
off and on again or the Alarm Silence button (11) is pressed.
COMPACT MAN-REV B 1/12/99, 9:47 AM59