Specifications
57
C or F. Selection of this button toggles the temperature display
between Celsius and Fahrenheit and produces an advisory that the
trend will be lost. Choosing Yes will cause the Monitor to accept
the new temperature label and erase the trend memory. Choosing
No will cause the Monitor to retain the existing temperature label
and the trend memory.
TREND WILL BE LOST
Yes
No
OK. Selection of this button returns the user to the Service mode
menu.
Note: Also on this panel is an indication of the current temperature
algorithm (3 Mins or 12 Mins), which tells the user whether the
temperature displayed by the Monitor will be representative of a
mercury-in-glass thermometer used for 3 minutes or 12 minutes,
respectively. The algorithm is set in the restricted service menu. The
initial factory default is 3 minutes.
Info. Selection of this button causes the most recent calibration
dates of the NIBP and temperature systems to be displayed.
Selection of OK returns the user to the Clinician menu.
OK
Last Calibration
NIBP: 14-Jul-1997
Temp: 14-Jul-1997
More.... Selection of this button displays the More... menu, which
allows the user to permanently change default mode settings.
Trend
Quik BP
Print
Set BP
Alarms Main
Using the Menu System
COMPACT MAN-REV B 1/12/99, 9:47 AM57