Specifications
56
Clinician Menu
Press Temp
More...
Silence Main
Info
Press. Selection of this button displays a panel for setting the
default BP target inflation pressure. Adjusting the default target
pressure will automatically update the current inflation target
pressure and will be used for the next reading. The range of
adjustment is 100 mmHg to 180 mmHg, and the setting is
retained when the Monitor is turned off.
The initial target pressure can be set between 100 and 180
mmHg in 5 mmHg increments. The factory default is 180 mmHg
for adults and 110 for neonates. When the target pressure is
changed, the next determination will use the new target inflation
value if no systolic is available. When adjusted under the Clinician
menu of the Service mode, the target pressure is adjusted
permanently.
TARGET PRESSURE
Default 100
Main
Temp. Selection of this button displays the temperature
submenu, which allows the user to choose the temperature
label. When C (Celsius) is selected, the °C indicator lights. When
F (Fahrenheit) is selected, the °F indicator lights.
OK
C or F
3 Mins
COMPACT MAN-REV B 1/12/99, 9:47 AM56