Specifications

50
History
Selection of this button causes the entire contents of the
trend memory to be printed. When selected, this button
temporarily changes to Cancel until the history has
completed printing.
Main
Selection of this button returns the user to the Main menu.
More... Menu
This menu is used to set the various operating modes of the
Monitor.
SpO
2
Config
Display
Service Main
Trend
SpO
2
Button (Models S and TS)
Selection of this button displays the SpO
2
mode menu,
which is used to set the SpO
2
response mode and pulse
tone volume.
1
Mode
Volume 5
Main
Mode
Three response modes are available. Each uses a different
averaging time while measuring to account for variations in
patient motion.
Mode 1 uses 5-7 second averaging; it is the most widely
used and the default mode.
Mode 2 uses 2-3 second averaging; it is useful for special
studies, but is sensitive to patient movement.
Mode 3 uses 10-15 second averaging and is least affected
by patient movement.
COMPACT MAN-REV B 1/12/99, 9:47 AM50