Specifications

48
Check
ALARM VOLUME
Main
5
Auto
Selection of this button updates the alarm limits on the LCD
relative to the current parameter values. Pressing this button
will automatically cancel any limit violation alarm that
becomes invalid as a result of a limit change. Alarm limits
are updated as follows:
Parameter Label
Systolic
Diastolic
MAP
Heart Rate
SpO
2
* If the reading plus the limit is greater than the valid range of measurement (e.g., SpO2 +5
is greater than 100% ), the valid range of measurement becomes the limits.
Label
SYS
DIA
MAP
BPM
SpO
2
Low Limit
SYS -30
DIA -30
MAP -30
BPM -30
SpO
2
-5*
High Limit
SYS +30
DIA +30
MAP +30
BPM +30
SpO
2
+5*
Notes
• In no case will the updated alarm limits be set beyond
the valid limits in the preceding table.
• If no values are available, the limits will remain
unchanged.
Main
Selection of this button returns the user to the Main menu.
Temp Button (Models T and TS)
Selection of this button causes a predictive temperature
measurement to be performed. This takes approximately 30
seconds to complete. Different predictive modes are used
for oral probes and rectal probes. The Monitor
automatically selects the correct mode for the probe type
used.