Specifications
47
mode (refer to “Press” in the “Using the Menu System”
section).
Main
Selection of this button returns the user to the Main menu.
Alarms Button
Selection of this button displays the Alarms menu. This
menu is used to adjust the violation limits for BP, Pulse Rate,
and SpO
2
. The values and ranges for these parameters are
not stored when the Monitor is turned off. The user may
edit the limits, but they are restored to the default values
each time the Monitor is switched on. To permanently
change the alarm limits, refer to “Alarms” under “Service
Button” in the “Using the Menu System” section.
Volume
SYS
Systolic High
Systolic Low
Diastolic High
Diastolic Low
MAP High
MAP Low
Heart Rate High
Heart Rate Low
Sp0
2
High
Sp0
2
Low
35 - 245
30 - 240
15 - 195
10 - 190
20 - 215
15 - 210
25 - 250
20 - 245
51 - 100
50 - 99
180
30
130
15
140
50
160
40
Off
90
DIA MAP BPM SpO
2
Auto
Main
180 130 140 160
OFF
30 15 50 40 90
Parameter Range Default
Volume
Selection of this button displays the alarm volume submenu.
The volume range is from 1 to 10, with 10 being the
loudest. The alarm volume is stored when the Monitor is
turned off and restored to the user’s preference each time
the Monitor is switched on. Selection of the Check button
allows the current volume setting to be heard. Selection of
the Main button returns the user to the Main menu.
Using the Menu System
COMPACT MAN-REV B 1/12/99, 9:47 AM47