Specifications

46
Set BP Button (UK: BP Mode)
Selection of this button displays the Auto and Manual BP menu.
Tgt Pressure
Main
Manual
Auto BP
180
Auto
Selection of this option starts an Auto BP determination. When
Auto Mode is selected, a number at the right of the Auto button
indicates the time interval between each reading. To change the
time interval, choose the box around the number and turn the
SelectKnob until the desired interval is reached. The interval can
be set between 1 and 90 minutes (1, 2, 3, 4, 5, 10, 15, 20, 25, 30,
45, 60, and 90 minutes). Press the SelectKnob to confirm the
setting. After power-off, the operating mode returns to the
default setting of Manual. The default setting of Manual can be
overridden to return to the previous user-selected setting (Auto or
Manual) by selecting Set BP under the Service menu.
Manual
Selection of this option allows the user to take a Manual BP
determination by simply pressing the Start/Stop BP key. After
power-off, the operating mode returns to the default setting of
Manual. The default setting of Manual can be overridden to
return to the previous user-selected setting (Auto or Manual) by
selecting Set BP under the Service menu.
Tgt Pressure
Selection of this option allows the user to set the BP target
inflation pressure. The initial target pressure can be set between
100 and 250 mmHg in 5 mmHg increments. The factory default
is 180 mmHg for adults and 110 for neonates. If the target
pressure is set to 140 mmHg under the Set BP or Clinical menu,
then the neonatal target pressure will be 110 mmHg. If the target
pressure is set between 100 and 140 mmHg, then that setting is the
target pressure that will be used. When the target pressure is
changed, the next determination will use the new target inflation
value if no systolic is available. Initial target pressure is restored to
the factory default setting after power-off. The target pressure can
be adjusted permanently in the Clinician menu of the Service
COMPACT MAN-REV B 1/12/99, 9:47 AM46