Installing the Temp Probe Holder
Attach the temperature probe holder to the side of the
DINAMAP Compact Monitor (Models T and TS) by aligning
the back and bottom edges of the holder and the Monitor
and pressing the holder firmly to the Monitor.
A distinct snap will sound when the Dual Lock
TM
fasteners
are properly engaged.
To remove the holder, place your fingers in the indentations
at the back of the Monitor and pull the holder away from
the side of the Monitor.
Switching the Monitor On and Off
To switch the DINAMAP Compact Monitor on, push the
power on/off switch (6) or click the SelectKnob (9).
As the Monitor powers up, it will run a short self-test
routine, which will flash all the indicator lights and then
beep the warning speaker. After a few seconds the system
will be ready for operation, as indicated by the appearance
of the main menu on the LCD (10).
To switch the Monitor off, push the power on/off switch (6)
again. This will terminate any measurements that may be in
progress and automatically deflate the cuff.
When the Monitor is operating on the internal battery only,
battery life is enhanced by the use of the sleep mode.
However, the Compact Monitor will not enter sleep mode if
an alarm is active. If no controls are used and no