Service manual

2048723-002D CARESCAPE V100 Vital Signs Monitor 5-7
Troubleshooting: Alarm code interpretation
Exergen-specific error codes
If the scanner is unable to take a temperature determination, or has a low
battery, the monitor will display theE-- error code and generate an audible
alarm. The scanner may also display additional indicators on the LED display.
Exergen LED display Condition Description
HI High Target > 43
°
C (110
°
F)
LO Low Target < 16
°
C (61
°
F)
HI A High Ambient > 40
°
C (104
°
F)
LO A Low Ambient < 16
°
C (60
°
F)
bAtt Low Battery Replace battery soon.
(blank display) Dead Battery Replace battery.
Err Processing Error Restart scanner. If unit is defective, it will have to be returned to the
manufacturer. Contact GE Technical Support.