Service manual

4-30 CARESCAPE V100 Vital Signs Monitor 2048723-002D
Maintenance: Test results form
Heart rate reading (bpm)
NIBP overpressure
E80’ displayed on Systolic display
Pump will not start
Overpressure alarm can be cleared
Display
All 7-Segment LEDs Light Correct Color
All Discrete LEDs Light, Correct Color
External DC detection
CHARGING indicator LED illuminated
Temperature test
Alaris
Temperature reading at 80.2° F 79.9° F 80.5° F
Temperature reading at 98.6° F 98.4° F 98.8° F
Temperature reading at 107.8° F 107.5° F 108.1° F
Exergen
Temperature reading of reference instrument
Temperature reading of instrument being verified
Difference between temperature reading of reference
instrument and instrument being verified
±0.2° C
(0.4° F)
SpO
2
Pulse Value Displayed
Saturation Value Displayed
Signal Strength Bar
--- displayed on SpO
2
display
Alarm is silenced, error display remains
Pulse Value
Saturation Value Displayed
Signal Strength Bar
Printer test
Printout is generated cleanly
Safety testing
Patient leakage
Fwd Pol, Neut Clsd, Gnd Clsd 100µA
Fwd Pol, Neut Open, Gnd Clsd
500µA
Fwd Pol, Neut Clsd, Gnd Open
500µA
Rev Pol, Neut Clsd, Gnd Open
500µA
Description Min Max Actual Pass-Fail-N/A