Service manual
2048723-002D CARESCAPE V100 Vital Signs Monitor 4-29
Maintenance: Test results form
Test results form
Description Min Max Actual Pass-Fail-N/A
NIBP Tests
Pneumatic leakage result (mmHg) 06
Pressure transducer verification
Pressure reading at 200mmHg, top display - Systolic 197 203
Pressure reading at 200mmHg, bottom display -
Diastolic
197 203
Pressure reading at 150mmHg, top display - Systolic 147 153
Pressure reading at 150mmHg, bottom display -
Diastolic
147 153
Pressure reading at 100mmHg, top display - Systolic 97 103
Pressure reading at 100mmHg, bottom display -
Diastolic
97 103
Pressure reading at 50mmHg, top display - Systolic 47 53
Pressure reading at 50mmHg, bottom display -
Diastolic
47 53
Overpressure verification
Overpressure threshold, Adult (mmHg) 305 325
Overpressure threshold, Neonate (mmHg) 150 165
Buttons
NIBP Determination Initiated
‘E80’ displayed on Systolic display
Audible alarm can be silenced
Alarm silence indicator (bell) is lit
Overpressure alarm can be cleared
Alarm button is functioning
NIBP determination
Systolic reading (mmHg)
Diastolic reading (mmHg)
MAP reading (mmHg)
Heart rate reading (bpm)
Systolic reading (mmHg)
Diastolic reading (mmHg)
MAP reading (mmHg)
Heart rate reading (bpm)
Systolic reading (mmHg)
Diastolic reading (mmHg)
MAP reading (mmHg)