2048723-002D CARESCAPE V100 Vital Signs Monitor 4-23
Maintenance: Safety testing
2. Firmly press the door to close it.
3. Press Print button.
4. Verify the printer outputs a record and print quality is good.
5. Confirm test results are recorded on the “Test results form” on page 4-29.
Safety testing
Electrical safety tests
Electrical safety tests provide a method of determining if potential electrical
health hazards to the patient or operator of the device exist.
Recommendations
Qualified personnel must perform all safety tests presented in this document:
Upon receipt of the monitor and its associated equipment).
Every 12 months thereafter (Planned/Preventive Maintenance). Refer to the
“Maintenance schedule” on page 4-2 for more information.
Each time the main enclosure is disassembled or a circuit board is removed,
tested, repaired, or replaced (Corrective Maintenance).
GE recommends that the qualified personnel performing the tests should
record the values of each required electrical safety test in the “Test results
form” on page 4-29. These instructions are intended for any component in
the system requiring electrical safety testing.