Service manual
4-22 CARESCAPE V100 Vital Signs Monitor 2048723-002D
Maintenance: Calibration procedures and tests
3. Remove the sensor from your finger to generate a ‘---’ (SpO
2
SENSOR OFF
FINGER) alarm and to sound the speaker.
4. Press the Silence button.
5. Verify the sound has stopped and the ‘---’ (SpO
2
SENSOR OFF FINGER) error
is cleared.
6. Re-apply the SpO
2
sensor to your finger.
7. Verify the unit displays a:
Pulse value
Saturation value
Signal Strength Bar Graph
8. Confirm all test results are recorded on the “Test results form” on page 4-29.
Printer output test
1. Load paper into the print mechanism.
NOTE
Use only printer paper from GE.
a. With the monitor powered on, turn it so that the side with the printer is
facing you.
b. While grasping the side of the monitor, lift the printer door open by
placing your thumb in the indented area and pulling. The printer door
will pop open.
c. Place the roll of paper into the compartment so that the end of the
paper comes off the right-side of the roll (paper is wound around the roll
clockwise). Place the roll of paper in the holding bracket that is
integrated in the door of the printer, making sure the paper extends out
of the printer cavity at least two inches.