4-18 CARESCAPE V100 Vital Signs Monitor 2048723-002D
Maintenance: Calibration procedures and tests
7. Press the Silence button again.
8. Verify the alarm condition is cleared from the Systolic window.
9. Confirm all test results are recorded on the “Test results form” on page 4-29.
LED tests
1. Power on the monitor.
2. During the power-up self-test verify all 7 segment LEDs and all discrete LEDs
illuminate and they are the correct color.
NOTE
In the following figure, the items with a solid border are segment LEDs,
and the items with a dotted border are discrete LEDs.
3. Repeat power up cycle until all LEDs are checked.
4. Confirm all test results are recorded on the “Test results form” on page 4-29.
External DC verification
1. Plug the power supply into the monitor.
2. Verify that the CHARGING indicator is illuminated.
3. Confirm all test results are recorded on the “Test results form” on page 4-29.
Temperature (perform if equipped with Temp module)
At installation, GE recommends performing a temperature check. Refer to
“Parameter level functional testing” - “Temperature” on page 4-11.