Service manual

2048723-002D CARESCAPE V100 Vital Signs Monitor 4-17
Maintenance: Calibration procedures and tests
Button testing
1. Disconnect the cuff/hose assembly and power on the unit.
2. Press Inflate/Stop button.
3. Verify a NIBP determination has been initiated.
4. Block pump port and verify E80 alarm.
5. Press Silence button, verify alarm has been cleared.
6. Verify that the alarm silence indicator (bell) is lit.
7. Press Alarm button several times, verify unit cycles through all alarm
settings (i.e., SYS, DIA, SpO2).
8. Confirm all test results are recorded on the “Test results form” on page 4-29.
9. Turn unit off.
NIBP functional test
NOTES
The NIBP function test does not test for measurement accuracy.
While an NIBP simulator device may be useful to verify that the monitor
responds to oscillometric pulsations, it should not be used as a basis for
assessing the accuracy of measurement. For more information, see
“Appropriate use of NIBP simulators” on page B-2.
1. Remove the calibration setup and attach an adult cuff and hose to a
simulator (be sure to select the correct cuff size).
2. Press Inflate/Stop button on monitor to begin a determination.
3. Record Systolic, Diastolic, MAP and heart rate from the monitor display.
4. Wait 1 minute, then press Cycle button to initiate a determination in Auto
NIBP mode.
5. Record Systolic, Diastolic, MAP and heart rate from the monitor display.
6. Wait 1 minute, then press Cycle button until stat is displayed in the min
window to initiate a determination in STAT mode.
7. Record Systolic, Diastolic, MAP and heart rate from the monitor display.
8. Press Inflate/Stop button, end STAT mode.
9. Confirm all test results are recorded on the “Test results form” on page 4-29.
NIBP overpressure verification
1. Remove the cuff/hose from the monitor, restrict airflow from cuff hose port.
2. Press Inflate/Stop to begin NIBP determination.
3. Verify E80is displayed on the Systolic window and audible alarm sounds.
4. Remove the air restriction.
5. Press Inflate/Stop and verify that the pump does not start.
6. Press the Silence button.