Service manual

2048723-002D CARESCAPE V100 Vital Signs Monitor 4-15
Maintenance: Calibration procedures and tests
The leakage rate is the difference between the first and second readings.
8. Record and verify the leakage rate on the “Test results form” on page 4-29.
9. Turn the monitor off.
Pressure transducer verification
NOTE
To enter Service Mode press and hold the Cycle button while pressing the
On/Off button.
1. Turn unit on and enter Service Mode.
2. The min window should display 0.
3. Open the valve on the inflation bulb and remove all pressure from the
system (manometer reads zero).
4. Press Cycle button and 1 should appear in the min window.
5. Use the inflation bulb to inflate the cuff, hose and pressure indicator setup to
200 mmHg.
6. Record and verify the pressure reading that appears in the Systolic window.
7. Record and verify the pressure reading that appears in the Diastolic
window.
8. Use the valve on the bulb to reduce pressure to 150 mmHg.
9. Record and verify the pressure reading that appears in the Systolic window.
10. Record and verify the pressure reading that appears in the Diastolic
window.
11. Use the valve on the bulb to reduce pressure to 100 mmHg.
12. Record and verify the pressure reading that appears in the Systolic window.
13. Record and verify the pressure reading that appears in the
Diastolic
window.
14. Use the valve on the bulb to reduce pressure to 50 mmHg.
15. Record and verify the pressure reading that appears in the Systolic window.
16. Record and verify the pressure reading that appears in the Diastolic
window.
17. Confirm all test results are recorded on the “Test results formon page 4-29.