Service manual

4-12 CARESCAPE V100 Vital Signs Monitor 2048723-002D
Maintenance: Parameter level functional testing
Exergen
1. Connect the Exergen temporal scanner’s modular plug to the host
communication port at the back of the monitor.
2. Take a temperature reading of yourself (i.e., not on a resistor simulator).
3. Verify that the temperature displays both on the Exergen temporal scanner
and in the Temperature window on the monitor.
Ohmeda, Nellcor, and Masimo SpO
2
technologies
The SpO
2
sensor is an assembly consisting of two parts: the sensor and the
extender cable.
1. Connect the cables prior to attaching to the monitor.
2. Verify that an SpO
2
reading is displayed within moments of attaching the
sensor to either a simulator or to your finger.