Service manual

2048723-002D CARESCAPE V100 Vital Signs Monitor 4-9
Maintenance: Long-term storage
6. Insert the battery into the compartment.
7. Replace the cover, help card tray, and screws. Insert the external DC power
converter plug into the external DC power socket and plug into an AC outlet.
NOTES
Error codeE00 appears (MEMORY LOST) alerting you that the user
settings (including alarm limits and inflation pressure) and date/time will
go back to default values and all entries in clinical history are erased.
For units with Ohmeda TruSignal technology, it is important to verify the
Line Frequency mode setting.
Configured settings and time/date will not be lost when reset due to
pressing and holding the On/Off button regardless of whether the DC
charger is attached.
8. Reset the date/time and applicable user settings.
9. Verify the configuration settings prior to returning the monitor to clinical use.
Exergen temporal scanner battery
NOTE
If the Exergen scanner is not used regularly, remove the battery to prevent
possible damage due to chemical leakage.
Replacing the Exergen temporal scanner battery
1. Disconnect the scanner cable from the monitor.
a. Loosen the two thumbscrews from the scanner’s modular plug.
b. Unplug the scanner cable from the monitor’s host communication port.