Service manual

2048723-002D CARESCAPE V100 Vital Signs Monitor 3-11
Installation: Configuring the monitor
3. To set the unit of measurement to:
°F move the switch up toward the probe cone
°C — move the switch away from the probe cone
4. Replace the cover, and tighten the screw.
Temperature display time configuration settings
User selects number of minutes temperature values are displayed.
NOTE
This procedure applies to both the Alaris probe and the Exergen scanner.
1. With the monitor off, press and hold the Menu button at the same time as
pressing the On/Off button until the display test completes.
2. Press the Menu button until tdt (temperature display time) appears in the
Pulse Rate window.
3. Use the +/- buttons to select the option.
4. To exit the configuration mode, turn the unit off. To continue with additional
configuration settings, press the Menu button.
Advanced configuration mode
Advanced Configuration mode is entered by holding the Menu button and the -
button simultaneously while powering on with the On/Off button.
For a few seconds immediately after power up in this mode, the Systolic and
Diastolic display windows will display the major and minor version codes. The
version codes are numbers that represent the letters of the English alphabet
which are designated to the currently loaded version of the monitor firmware
(e.g., 1 indicates A, 2 indicates B, etc.).
After a moment, this version information is cleared, and the monitor displays the
first page of configuration mode which simply displays ACF in the Systolic
display window indicating that the monitor is in advanced configuration mode.
F/C switch
°F
°C