Service manual

2048723-002D CARESCAPE V100 Vital Signs Monitor 3-7
Installation: Configuring the monitor
Setting the date and time – To set the date and time on the monitor, you must
access the configuration mode. Press Menu to skip the default settings that do
not require changes. Refer to the table above.
NOTE
While in configuration mode, all entries stored in the clinical history are
erased when the time and/or date is changed.
Procedures
1. Press the Menu button to move from one setting to another. Use the +/-
buttons to increment or decrement the setting.
NOTE
For the date and time to be saved, you must advance the menu through
the minute setting.
2. To exit the configuration mode, press the On/Off button.
3. To continue with other changes, press the Menu button. will appear in
the Systolic window. To change parameter settings, press the Menu button
SpO
2
sat
(Masimo only)
SpO
2
0 or 1 Fast Sat Mode
0=Off
1=On*
SpO
2
sensitivity
(Masimo only)
SpO
2
1, 2, 3 1= Low Perfusion-Maximized,
2= Low Perfusion-Default,
3= for engaging Adaptive Probe
Off Technology algorithm*
Temperature units Temperature blank C or F indicator illuminated (Alaris
only)
Temperature display
time
Temperature 2 or 5 User selects number of minutes
temperature values are
displayed.
Year Systolic XXX (numeric) use + and - keys to change
Month MAP/Cuff XXX (numeric) use + and - keys to change
Day Diastolic XXX (numeric) use + and - keys to change
Hour min XXX (numeric) use + and - keys to change
Minute min XXX (numeric) use + and - keys to change
Mode
(when main screen is
active)
Systolic blank indicates configuration mode
NOTE: Refer to the “SpO
2
section of operator’s manual for detailed descriptions of the different user selectable SpO
2
settings.
Setting Window LED display
Pulse Rate
window
display
Comment