3-6 CARESCAPE V100 Vital Signs Monitor 2048723-002D
Installation: Configuring the monitor
These displays appear only during the first part of the power up sequence and
are not selectable and cannot be changed. After a moment, this version
information is cleared, and the monitor displays the first page of configuration
mode which simply displays in the Systolic window.
Pressing the Menu button cycles through all the configuration option pages.
After all options pages have been displayed, the display returns to the first
configuration mode page (displaying ). You can use the + and - buttons to
make changes to settings. After making changes, simply cycle the power to
return to normal operation (clinical) mode. Changes are automatically retained.
NOTES
If a parameter is not enabled, its options are not displayed.
Menu selections for SpO
2
settings are different depending upon the SpO
2
technology your monitor contains.
The Menu selections appear in the following order. Refer to the manual section
for settings options.
Display Window
Major software revision Systolic
Minor software revision Diastolic
Type of NIBP technology min
Temperature type technology Temperature
Setting Window LED display
Pulse Rate
window
display
Comment
Inflate pressure
(adult/ped)
Systolic XXX (numeric) XXX (numeric) ADULT indicator illuminated,
INFLATE PRESSURE indicator
flashing
Inflate pressure
(neonate)
Systolic XXX (numeric) XXX (numeric) NEONATE indicator illuminated,
INFLATE PRESSURE indicator
flashing
Line frequency mode
(Ohmeda TruSignal only)
SpO
2
50 or 60 AC line frequency
SpO
2
mode
(Nellcor only)
SpO
2
1 or 2 User selects the averaging
technique*
1=Normal Response
2= Fast Response
SpO
2
sat
(Nellcor only)
SpO
2
0, 10, 25, 50, 100 User selects the SMART Sat
tolerance level *
SpO
2
mode
(Masimo only)
SpO
2
4, 6, 8, 10, 12, 14, 16 User selects the number of
seconds over which data is
averaged 4 to 16*