Service manual

3-4 CARESCAPE V100 Vital Signs Monitor 2048723-002D
Installation: Powering the monitor
When about 5 minutes of main battery charge remains:
The low-priority BATTERY LOW alarm escalates to a high-priority BATTERY LOW
alarm.
The BATTERY LOW indicator flashes.
Any NIBP determination in progress at the time of the alarm escalation is
allowed to finish.
Any Stat mode cycle that was initiated before the alarm escalation is
allowed to finish.
The user is not able to initiate:
any new NIBP determinations of any type
any printouts
NOTE
At this time, it is highly recommended to plug the monitor into external DC
power.
When 5 minutes of main battery charge expires:
After 5 minutes of high-priority BATTERY LOW alarm, the monitor enters a
battery low shutdown.
No error code is displayed.
The BATTERY LOW indicator flashes.
The monitor alarms for 2.5 minutes, then shuts down completely.
NOTE
You must plug the monitor into DC power before resuming monitoring.
After plugging the monitor into DC power:
The BATTERY LOW indicator (when the monitor is on) and CHARGING
indicator illuminate.
The BATTERY LOW indicator turns off when the battery level reaches a
sufficient charge level to operate without the BATTERY LOW alarm active.
E13 BATTERY LOW
At any time while the high-priority BATTERY LOW alarm is active, certain actions
can trigger the ‘E13 BATTERY LOW alarm: any attempt to start an NIBP
determination or a printout. This alarm is giving you additional warning that the
battery charge is critically low.
NOTE
At this time, it is highly recommended to plug the monitor into external DC
power.
The ‘E13 error code appears in the min window.
The BATTERY LOW indicator flashes.
This alarm can be acknowledged by pressing the Silence button.
The user is not able to initiate:
any new NIBP determinations of any type
any printouts