Service manual

2048723-002D CARESCAPE V100 Vital Signs Monitor 2-3
Equipment overview: Front controls and connectors
Front controls and connectors
1. Silence button: mutes audible alarms. Any other active alarm that can be
acknowledged is also cleared and the alarm condition is reset whenever this
key is pressed. When pressed, the alarm silence indicator (bell) lights solid
red to indicate that audible alarms have been silenced for 2 minutes. Alarm
silence can be cancelled by pressing the Silence button again.
2. Alarms button: used to view or adjust parameter alarm limit settings.
3. +/- buttons (Plus/Minus): used when you are in the following modes: limit,
menu, cycle, and history.
When you are in limit or menu setting, pressing the +/- button increases
and decreases an adjustable setting.
When you are in cycle or history mode, pressing the +/- buttons
displays the next or previous cycle selection or entry in the history list,
respectively.
When you reach the beginning or ending of a list, a negative key-click
sounds.
4. Menu button: accesses menu settings that can be adjusted: INFLATE
PRESSURE (ADULT and NEONATE), ALARM VOLUME, and PULSE VOLUME.
(Refer to “Operating modes” on page 3-5 for a description of clinical mode.)
NOTE
ADULT indicator encompasses both adult and pediatric patients.
5. SpO
2
sensor connector: attach SpO
2
cables here.
6. NIBP connector: attach NIBP cuff hoses here.
7. Inflate/Stop button: starts a manual NIBP determination or stop any NIBP
determination.
8. Temperature probe holster: stores Alaris temperature probe.