Electromagnetic compatibility (EMC): Electromagnetic compatibility (EMC): CARESCAPE V100 monitor
Electromagnetic compatibility (EMC): CARESCAPE V100
monitor
Changes or modifications to this system not expressly approved by GE
Healthcare can cause EMC issues with this or other equipment. This system is
designed and tested to comply with applicable regulation regarding EMC and
must be installed and put into service according to the EMC information stated
in this appendix.
CAUTION
Use of portable phones or other radio frequency (RF) emitting
equipment near the system may cause unexpected or adverse
operation.
CAUTION
The equipment or system should not be used adjacent to, or
stacked with, other equipment. If adjacent or stacked use is
necessary, the equipment or system should be tested to verify
normal operation in the configuration in which it is being used.
Guidance and manufacturer’s declaration – electromagnetic emissions
The CARESCAPE V100 vital signs monitor is intended for use in the
electromagnetic environment specified below. It is the responsibility of the
customer or user to assure that the CARESCAPE V100 vital signs monitor is used
in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF Emissions
EN 55011
Group 1 The equipment uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF Emissions
EN 55011
Class B
Harmonic Emissions
IEC 61000-3-2
Class A The equipment is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
Complies