Service manual

Technical specifications and default settings: Specifications
2048723-002D CARESCAPE V100 Vital Signs Monitor A-15
Battery
Monitor - main battery
Exergen temporal scanner
Battery specifications
Capacity 6V; 3.3 Ahr sealed lead-acid battery
Battery life 5 hours with SpO
2
technology with a usage scenario of:
NIBP determinations every 5 minutes with adult cuff
Printout after every determination
SpO
2
parameter active at 60 bpm
Temperature parameter active in monitor mode
Up to 11.5 hours without SpO
2
technology with a usage scenario of:
NIBP determinations every 15 minutes
Temperature parameter active
Charge time Approx. 5 hours from full discharge when the monitor is off
Approx. 8 hours when the monitor on
Battery specifications
Capacity (1) 9V alkaline
Battery life Approx. 15,000 readings
(When scanning for 5 seconds and reading the temperature display for
2 seconds before turning thermometer off.)