Service manual

A-14 CARESCAPE V100 Vital Signs Monitor 2048723-002D
Technical specifications and default settings: Specifications
Exergen
Measurement Mode Max: 43°C; 110°F
Min: 16°C; 61°F
Accuracy ±0.1°C; ±0.2°F
meets EN 12470-5
Predictive Mode Not applicable
Monitor Mode Not applicable
Operating environment 16° to 40°C; 60° to 104°F (ambient)
Arterial head balance range for body
temperature*
34.5° to 43°C; 94° to 110°F
Resolution 0.1°C or 0.1°F
Response time approx. 0.04 seconds, typical
*
Automatically applied when temperature is within normal body temperature range, otherwise reads surface temperature.
Temperature specifications