Technical specifications and default settings: Specifications
2048723-002D CARESCAPE V100 Vital Signs Monitor A-13
Temperature
WARNING
Alaris: Use only recommended probes and probe covers. The
size, shape, and thermal characteristics of the probe covers
can affect the performance of the instrument. Inaccurate
readings may occur unless recommended probes and probe
covers are used. Visually inspect the probe prior to use to be
sure it is defect free. Please refer to the Accessories list
provided with your monitor.
Temperature specifications
Units of measure °Fahrenheit (F) or °Celsius (C)
Range
Alaris
Predictive Mode
Turbo Temp Max: 41.1°C; 106.0°F
Min: 35.6°C; 96.0°F
Tri-Site Max: 41.1°C; 106.0°F
Min: 35.0°C; 95.0°F
Monitor Mode
Turbo Temp Max: 42.1°C; 107.9°F
Min: 26.7°C; 80.0°F
Tri-Site Max: 42.1°C; 107.9°F
Min: 26.7°C; 80.0°F
Turbo Temp and Tri-Site
Continuous (monitor) mode accuracy
±0.1°C; ±0.2°F
when tested in a calibrated liquid bath;
meets ASTM E1112, Table 1, in range
specified
NOTE:
If large changes occur in the ambient
temperature, the temperature system
can be recalibrated by cycling the
monitor’s power using the On/Off
button.
Determination time
Turbo Temp
Oral or rectal As fast as 7 seconds
Tri-Site
Oral
Rectal
Axillary
As fast as 12 seconds
As fast as 11 seconds
As fast as 13 seconds