Service manual
A-12 CARESCAPE V100 Vital Signs Monitor 2048723-002D
Technical specifications and default settings: Specifications
LNCS DC-I ± 2 digits without motion
LNCS DC-IP ± 2 digits without motion
LNCS Adult Adtx ± 2 digits without motion
LNCS Ped Pdtx ± 2 digits without motion
LNCS Infant-L ± 2 digits without motion
LNCS Neo PT-L ± 3 digits without motion
Resolution
Saturation (% SpO
2
)1%
Pulse rate (bpm) 1
Low perfusion performance
0.02% Pulse amplitude
and % transmission >5%
Saturation (% SpO
2
) ±2 digits
Pulse rate ±3 digits
Interfering substances Carboxyhemoglobin may erroneously increase
readings. The level of increase is approximately
equal to the amount of carboxyhemoglobin
present. Dyes, or any substance containing dyes,
that change usual arterial pigmentation may
cause erroneous readings.
Sensor light source
Wavelength** Infrared: 905 nm (nominal)
Red: 660 nm (nominal)
Power dissipation Infrared: 22.5 mW (max)
Red: 27.5 mW(max)
*Masimo SET Technology with LNOP Blue sensors have been validated for no motion accuracy in human blood studies on neonatal, infant, and
pediatric patients with congenital, cyanotic cardiac lesions in the range of 60% to 100% SpO
2
against a laboratory CO-oximeter. This variation
equals plus or minus one standard deviation, which encompasses 68% of the population.
** Information about wavelength range can be especially useful to clinicians.
Specifications: Masimo sensor accuracy