Service manual

Technical specifications and default settings: Specifications
2048723-002D CARESCAPE V100 Vital Signs Monitor A-11
Specifications: Masimo sensor accuracy
Sensor model
SpO
2
range
70% to 100%
LNOP
LNOP ADT ± 2 digits without motion
LNOP NEO ± 3 digits without motion
LNOP NEO-L
Foot
Finger
± 3 digits without motion
± 2 digits without motion
LNOP NEO PT-L ± 3 digits without motion
LNOP Adtx ± 2 digits without motion
LNOP Pdtx ± 2 digits without motion
LNOP DCI ± 2 digits without motion
LNOP DCIP ± 2 digits without motion
LNOP Hi Fi-Neo/adult
Foot
Finger
± 3 digits without motion
± 2 digits without motion
LNOP Hi Fi-Infant/Ped ± 2 digits
LNOP Blue Infant Thumb/Toe* ± 3 digits (for 80-100) without motion
± 4 digits (for 60-80) without motion
± 3.3 digits (for 70-100) without motion
LNOP YI Multi-Site
Foot/hand
Finger/toe
± 3 digits without motion
± 2 digits without motion
LNOP DC-195 ± 2 digits without motion
LNOP TC-I ± 3.5 digits without motion
LNCS TCI ± 3.5 digits without motion