Technical specifications and default settings: Specifications
Masimo SpO
2
WARNING
Use only Masimo oximetry sensor for SpO
2
measurements.
Other oxygen transducers (sensors) may cause improper SpO
2
performance.
WARNING
Monitors identified as having Masimo SET technology are
compatible only with Masimo SET sensors and cables.
Masimo SpO
2
specifications
Measurement range
SpO
2
1 to 100%
Pulse rate 25 to 240 bpm
Perfusion range 0.02 to 20%
Accuracy and motion tolerance
Saturation
Without motion - adult/pediatric* 70 to 100% ± 2 digits
Without motion - neonate* 70 to 100% ± 3 digits
With motion - adult/pediatric/neo**
70 to 100% ± 3 digits
Low perfusion‡ 70 to 100% ± 2 digits
0 to 69% unspecified
Pulse rate
Without motion 25 to 240 bpm ±3 digits
With motion normal physiologic range
25 to 240 bpm ±5 digits
* The Masimo SET SpO
2
parameter with LNOP-Adt sensors has been validated for no motion accuracy in human blood studies on healthy adult
volunteers in induced hypoxia studies in the range of 70-100% SpO
2
against a laboratory CO-oximeter and ECG monitor. This variation equals plus
or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population.
**The Masimo
SET
SpO
2
parameter with LNOP-Adt sensors has been validated for motion accuracy in human blood studies on healthy adult volunteers
in induced hypoxia studies while performing rubbing and tapping motions at 2 to 4 Hz at an amplitude of 1 to 2 cm and a nonrepetitive motion between
1 to 5 Hz at an amplitude of 2 to 3 cm in induced hypoxia studies in the range of 70-100% SpO
2
against a laboratory CO-oximeter and ECG monitor. This
variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population.
†The Masimo
SET
SpO
2
parameter with LNOP-Neo Pt sensors has been validated for neonatal motion accuracy in human blood studies on neonates
while moving the neonate’s foot at 2 to 4 cm against a laboratory CO-oximeter and ECG monitor. This variation equals plus or minus one standard
deviation. Plus or minus one standard deviation encompasses 68% of the population.
‡The Masimo
SET
SpO
2
parameter has been validated for low perfusion accuracy in bench-top testing against a Biotek Index 2 simulator and
Masimo’s simulator with signal strengths of greater than 0.02% and a % transmission of greater than 5% for saturations ranging from 70 to 100%.
This variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population.