Service manual
Technical specifications and default settings: Specifications
2048723-002D CARESCAPE V100 Vital Signs Monitor A-9
D-YS & D-YSE ± 3.5 digits
D-YS & D-YSPD ± 3.5 digits
DS-100A ± 3 digits
OXI-A/N (adult) ± 3 digits
OXI-A/N (neonate) ± 4 digits
OXI-P/I ± 3 digits
Neonatal sensor accuracy When sensors are used on neonatal subjects as
recommended, the specified accuracy range is increased
by ±1 digit, as compared to adult usage, to account for the
theoretical effect on oximeter measurements of fetal
hemoglobin in neonatal blood. For example, MAX-N
accuracy on neonates is ±3 digits, rather than ±2 digits.
Sensor light source
Wavelength*** Infrared: 890 nm (nominal)
Red: 660 nm (nominal)
Power dissipation Infrared: 22.5 mW (max)
Red: 30 mW (max)
*
The MAX-N, D-YS, OXI-A/N, and OxiCliq N were tested on patients >40 kg.
**
The accuracy specification has been determined between saturations of 80%-100%.
*** Information about wavelength range can be especially useful to clinicians.
Specifications: Nellcor sensor accuracy