Service manual
Technical specifications and default settings: Specifications
2048723-002D CARESCAPE V100 Vital Signs Monitor A-7
Nellcor SpO
2
WARNING
Nellcor-labeled monitors are only compatible with Nellcor
Oximax sensors and cables.
Nellcor SpO2 specifications
Measurement range
SpO
2
1 to 100%
Pulse rate 20 to 250 bpm
Perfusion range 0.03 to 20%
Accuracy
Saturation
Adult* 70 to 100% ±2 digits
Neonate* 70 to 100% ±3 digits
Low perfusion** 70 to 100% ±2 digits
Pulse Rate
Adult and neonate 20 to 250 bpm ±3 digits
Low perfusion** 20 to 250 bpm ±3 digits
*Adult specifications are shown for OxiMax MAX-A and MAX-N sensors with the N-600. Saturation accuracy will vary by sensor type. This variation
equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population. Accuracy is based on deep
hypoxia studies on healthy adult volunteer subjects. Arterial blood samples were analyzed simultaneously on multiple CO-oximeters.
**Applicability: OxiMax MAX-A, MAX-AL, MAX-P, MAX-I, and MAX-N sensors.