Service manual
A-6 CARESCAPE V100 Vital Signs Monitor 2048723-002D
Technical specifications and default settings: Specifications
Specifications: Ohmeda sensor accuracy
Sensor model
SpO
2
range
70% to 100%
OxyTip++
TS-F-D ±2 digits without motion
TS-E-D ±3 digits without motion
TS-W-D ±2 digits without motion
TS-SE-3 ±2 digits without motion
TS-AF-10 ±2 digits without motion
TS-AF-25 ±2 digits without motion
TS-F4-GE ±2 digits without motion
TS-F2-GE ±2 digits without motion
TS-E4-GE ±3 digits without motion
TS-E2-GE ±3 digits without motion
Specifications: Ohmeda sensor accuracy
Sensor model
SpO
2
range
70% to 100%
SpO
2
range
90% to 100%
SpO
2
range
80% to 90%
SpO
2
range
70% to 80%
SpO
2
range
below 70%
TS-SA4-GE
±2 digits ±1 digits ±2 digits ±3 digits unspecified
TS-SA-D
Specifications: Ohmeda sensor accuracy
Sensor light source
Wavelength* Infrared: 930 to 950 nm (nominal)
Red 650 to 670 nm (nominal)
Average power < 1 mW
* Information about wavelength range can be especially useful to clinicians.